1:Adjustable height
2:Credible quality
3:CE/ISO/Certificate of conformity
4:Decades years manufacture experience
Main Use
and Applicable Scope:
The product is applicable for the body
temperature maintaining, emergency operation
and intensive care of new-born baby, sick
and weak infant and premature infant.
3.2 Normal Working
Environment
a) Power voltage: (220±22)V (50±1)Hz;
b) Ambient temperature: 18~30°C, with the
control temperature being at least 3°C
higher than the ambient temperature
3.3 Product
composition: temperature controller,
radiation source, bed frame, spot light and
frame, etc.
3.4 Heater:
radiation-type far IR heating element.
3.5 Normal Working
Environment:
a) Ambient temperature: 18~30°C, with the
control temperature being at least 3°C
higher than the ambient temperature;
b) Relative humidity: 30~75%;
c) Atmospheric pressure: 70~106Kpa;
d) Ambient air speed: < 0.3 m/s,
e) Placed in an arena with relatively static
air flow and without direct sunlight and
other heat sources.
f) Power voltage: 220±22V, 50±1Hz
g) Consumed power: ≤1000VA
3.6 Basic
Parameters
1) Illumination of spot light at the bed
center: ≥100lx
2) Constant temperature time: not less than
2h
3) Temperature uniformity: ±2°C with the
baby bed at the horizontal level:
4)Temp-controlled
mode:
Mode for infant temp-control: alarming when
the temperature deviating by +1.5°C and
-2.0°C after constant temperature is set.
Manual temp-control mode: the heater outputs
at fixed energy level and at max output
energy level.
5) Measuring range of skin temperature:
25.0°C~42.0°C
6) Setting range of skin temperature:
25.0°C~37.0°C
7) Operating time of manual temp-control:
set as 15min